100% Original Endotoxin free glass tube - Endotoxin Challenge Vials (Endotoxin Indicator) – Bioendo
100% Original Endotoxin free glass tube - Endotoxin Challenge Vials (Endotoxin Indicator) – Bioendo Detail:
Endotoxin Challenge Vials (Endotoxin Indicator)
1. Product Information
The Endotoxin Challenge Vial (ECV,Endotoxin Indicator) is used in the validation of dry heat depyrogenation cycles. Endotoxin Challenge Vials are placed in the cold spots of the dry heat oven. After cycle completion, the log reduction in endotoxin levels can be determined by comparing the endotoxin levels in the baked vs not baked Endotoxin Indicators. The Endotoxin Challenge vials are designed to indicate a minimum 3-log reduction in endotoxin content when tested using gel clot endotoxin assay kit, kinetic turbidimetric endotoxin assay kit or chromogenic Lyophilized Amebocyte Lysate assay kits. Endotoxin Challenge Vial contains 1000 to 10000EU Endotoxins level in a glass vial. ECV1250V and ECV2500V are sealed in vials that configurate the heat resistant stoppers and heat resistant labels, which could be placed in dry-heat oven or dry-heat tunnel without removal of stoppers and caps(lid).
Validation of dry heat sterilization cycle(s) is required by regulations of ANSI, AAMI, ISO, USP and FDA to ensure that all items that are required to be sterile or endotoxin-free are able to consistently and reliably be sterilized to reduce the chance of introducing or spreading an infectious micro-organism or pyrogens.
Follow the Regulations requirement, use Bioendo ECV is able to operate the endotoxin free validation or pyrogen free validation. More important information is Bioendo ECV no need rubber lid(cap) removal in the heating operation.
2. Product Parameter
The potency: 1000 to 10000 EU per vial and 2000 to 10000 EU in a vial.
3. Product Features and Application
In vial with heat resistant stopper (tolerable high temperature up to 360 Celsius degree) and heat resistant label.
In the procedure of baking or drying, ECV vial’s lid no need to remove.
Without excipient.
Catalog No. |
Endotoxins Level (EU/vial) |
Package |
ECV1250V |
1000-10000 |
In Vial, 10 Vials/Pack |
ECV1250VR |
1000-10000 |
ISO 2R vial, 10 Vials/Pack |
ECV2500V |
2000-10000 |
In Vial, 10 Vials/Pack |
ECV2500VR |
2000-10000 |
ISO 2R vial, 10 Vials/Pack |
ECV100000V |
50000-200000 |
In Vial, 10 Vials/Pack |
Endotoxin challenge vial, in the operation procedures, the lysate reagent has to match the ECV kit, and no need remove the lit during the baking procedure.
Product detail pictures:
Related Product Guide:
Sticking for the principle of "Super Quality, Satisfactory service" ,We have been striving for being a superb small business partner of you for 100% Original Endotoxin free glass tube - Endotoxin Challenge Vials (Endotoxin Indicator) – Bioendo , The product will supply to all over the world, such as: Munich, kazakhstan, Kuala Lumpur, With well educated, innovative and energetic staff, we are responsible for all elements of research, design, manufacture, sale and distribution. By studying and developing new techniques, we are not only following but also leading fashion industry. We listen attentively to the feedback from our customers and provide instant replies. You will instantly feel our professional and attentive service.
By Lillian from Orlando - 2018.09.21 11:01
Products and services are very good, our leader is very satisfied with this procurement, it is better than we expected,
By Eve from Guatemala - 2017.01.28 19:59