We have a professional, efficiency team to provide quality service for our customer. We always follow the tenet of customer-oriented, details-focused for Bacterial Endotoxin Test Procedure, Endotoxin Lipopolysaccharide, limulus lysate assay, Welcome any of one's inquiries and concerns for our items, we glance forward to creating a long-term business enterprise marriage along with you while in the around long run. call us today.
Well-designed Bacterial Endotoxin Calculation - Endotoxin Challenge Vials (Endotoxin Indicator) – Bioendo Detail:

 Endotoxin Challenge Vials (Endotoxin Indicator)

 

1. Product Information

The Endotoxin Challenge Vial (ECV,Endotoxin Indicator) is used in the validation of dry heat depyrogenation cycles. Endotoxin Challenge Vials are placed in the cold spots of the dry heat oven. After cycle completion, the log reduction in endotoxin levels can be determined by comparing the endotoxin levels in the baked vs not baked Endotoxin Indicators. The Endotoxin Challenge vials are designed to indicate a minimum 3-log reduction in endotoxin content when tested using gel clot endotoxin assay kit, kinetic turbidimetric endotoxin assay kit or chromogenic Lyophilized Amebocyte Lysate assay kits. Endotoxin Challenge Vial contains 1000 to 10000EU Endotoxins level in a glass vial. ECV1250V and ECV2500V are sealed in vials that configurate the heat resistant stoppers and heat resistant labels, which could be placed in dry-heat oven or dry-heat tunnel without removal of stoppers and caps(lid).

Validation of dry heat sterilization cycle(s) is required by regulations of ANSI, AAMI, ISO, USP and FDA to ensure that all items that are required to be sterile or endotoxin-free are able to consistently and reliably be sterilized to reduce the chance of introducing or spreading an infectious micro-organism or pyrogens.

Follow the Regulations requirement, use Bioendo ECV is able to operate the endotoxin free validation or pyrogen free validation. More important information is Bioendo ECV no need rubber lid(cap) removal in the heating operation.

 

2. Product Parameter

The potency: 1000 to 10000 EU per vial and 2000 to 10000 EU in a vial.  

 

3. Product Features and Application

In vial with heat resistant stopper (tolerable high temperature up to 360 Celsius degree) and heat resistant label. 

In the procedure of baking or drying, ECV vial’s lid no need to remove. 

Without excipient.  

Endotoxin Challenge Vial  (10ml glass bottle and 2ml of ISO standard 2R glass bottle)

Catalog No.

Endotoxins Level (EU/vial)

Package

ECV1250V

1000-10000

In Vial, 10 Vials/Pack

ECV1250VR

 1000-10000

ISO 2R vial, 10 Vials/Pack

ECV2500V

2000-10000

In Vial, 10 Vials/Pack

ECV2500VR

 2000-10000

ISO 2R vial, 10 Vials/Pack

ECV100000V

50000-200000

In Vial, 10 Vials/Pack

How is the features of Bioendo’s Endotoxin Challenge Vails (Endotoxin Indicator)?

Endotoxin challenge vial, in the operation procedures, the lysate reagent has to match the ECV kit, and no need remove the lit during the baking procedure.


Product detail pictures:

Well-designed Bacterial Endotoxin Calculation - Endotoxin Challenge Vials (Endotoxin Indicator) – Bioendo detail pictures

Well-designed Bacterial Endotoxin Calculation - Endotoxin Challenge Vials (Endotoxin Indicator) – Bioendo detail pictures

Well-designed Bacterial Endotoxin Calculation - Endotoxin Challenge Vials (Endotoxin Indicator) – Bioendo detail pictures

Well-designed Bacterial Endotoxin Calculation - Endotoxin Challenge Vials (Endotoxin Indicator) – Bioendo detail pictures


Related Product Guide:

To meet the customers' over-expected gratification , we have our robust crew to offer our best over-all support which includes marketing, income, coming up with, production, excellent managing, packing, warehousing and logistics for Well-designed Bacterial Endotoxin Calculation - Endotoxin Challenge Vials (Endotoxin Indicator) – Bioendo , The product will supply to all over the world, such as: Slovakia, Costa rica, Ecuador, Our staffs are rich in experience and trained strictly, with qualified knowledge, with energy and always respect their customers as the No. 1, and promise to do their best to deliver the effective and individual service for customers. The Company pays attention to maintaining and developing the long-term cooperation relationship with the customers. We promise, as your ideal partner, we'll develop a bright future and enjoy the satisfying fruit together with you, with persisting zeal, endless energy and forward spirit.
  • This is a very professional wholesaler, we always come to their company for procurement, good quality and cheap.
    5 Stars By Andrea from panama - 2018.11.22 12:28
    Hope that the company could stick to the enterprise spirit of "Quality, Efficiency, Innovation and Integrity", it will be better and better in the future.
    5 Stars By Judy from Brisbane - 2018.07.27 12:26

    Leave Your Messages

    Related products

    • OEM Manufacturer BET Test Procedure - Rapid Gel Clot 10 Samples Kit – Bioendo

      OEM Manufacturer BET Test Procedure - Rapid Ge...

      Rapid Gel Clot Endotoxin Assay Kit (10 Samples Kit) 1. Product Information Bioendo Rapid Gel Clot Endotoxin Assay Kit is designed to rapidly quantify endotoxin in water or dialysate. Generally, RG kit’s result could be gained within 30 minutes. Under the guidance of detecting endotoxin in water or dialysate quickly, endotoxin detection with Bioendo Rapid Gel Clot Endotoxin Assay Kit does not need the multi steps’ dilution of Control Standard Endotoxin and test samples. Operation procedure...

    • Factory Free sample Bacterial Endotoxin - Beta-glucans Blocker for Blocking Beta Glucan Pathway – Bioendo

      Factory Free sample Bacterial Endotoxin - Beta...

      Beta-glucans Blocker for Blocking Beta Glucan Pathway 1. Product information There are two pathway in limulus amoebocyte lysate lal reagent, factor C pathway is specific to endotoxin and factor G pathway is specific to(1,3)- β-D-Glucans. If the test sample contains β-1,3-Glucans, the limulus test (Endotoxin test ) will have interferences. β-G-Blocker blocks the reactivity of LAL to β-1,3-Glucans,conferring increased endotoxin specificity to the LAL test. If the test samples containsβ-1,3-Gl...

    • OEM/ODM China Dry heat sterilizer validation by ECV - Nymph X Water Purification System – Bioendo

      OEM/ODM China Dry heat sterilizer validation by...

      NYMPHX Nymph X Water Purification System   Nymph X Water Purification System could transfer tap water into pure water and ultrapure water. Equipped with a pressure stabilizer and low flow-rate operating mode, the system can adapt to various source conditions without any extra pre-treatment necessary. The system also supports multiple water dispensing modes, and the precision of quantitative and quality-controlled dispense of water could reach ±1%. Meanwhile, comprehensive monitoring could be...

    • China wholesale endotoxin test - Depyrogenated Sample Bottles ( Endotoxin Free ) – Bioendo

      China wholesale endotoxin test - Depyrogenated...

      Depyrogenated Sample Container 1. Product information We offer various of low endotoxin, pyrogen free accessories products,includes Water for Bacterial Endotoxins Test, pyrogen-free test tubes, pyrogen-free pipettor tips, microplates and sample bottles for your conveniences. High quality depyrogenated low endotoxin pyrogen free products insure the success of your experiments. Depyrogenated (Endotoxin Free) Sample container(endotoxin free bottle, pyrogen free bottle, pyrogen free sample bot...

    • Factory selling Bacterial Endotoxin Limit For API - Endotoxin-free Glass Test Tubes – Bioendo

      Factory selling Bacterial Endotoxin Limit For A...

      Endotoxin-free Glass Test Tubes   1. Product Information Endotoxin-free glass test tubes contain less than 0.005EU/ml endotoxin. Catalog number T107505 and T107540  are recommended for use as reaction tubes in gel clot and end-point chromogenic assays. Catalog number T1310018 and T1310005 is recommended for dilution of endotoxin standards and test samples. T1050005C is a special designed short endotoxin reaction tube that allows the pipette tips reaches the tube bottom. T087565 is used ...

    • Hot New Products LAL reagent sensitivity - Rapid Gel Clot 10 Samples Kit – Bioendo

      Hot New Products LAL reagent sensitivity - Rap...

      Rapid Gel Clot Endotoxin Assay Kit (10 Samples Kit) 1. Product Information Bioendo Rapid Gel Clot Endotoxin Assay Kit is designed to rapidly quantify endotoxin in water or dialysate. Generally, RG kit’s result could be gained within 30 minutes. Under the guidance of detecting endotoxin in water or dialysate quickly, endotoxin detection with Bioendo Rapid Gel Clot Endotoxin Assay Kit does not need the multi steps’ dilution of Control Standard Endotoxin and test samples. Operation procedure...